Endress+Hauser LPGmass D8EB Coriolis flowmeter

The refueling and distribution application flowmeter with easy system integration.

Free shipping
Credit Card

Overview

The LPGmass is specially designed for flow measurement of LPG for dispensing and truck unloading. It combines an integrated temperature measurement with intelligent conversion functions, providing volume correction directly on site. LPGmass will be the preferred choice for system integrators, skid builders and equipment manufacturers.

Benefits

  • Excellent operational safety – reliable under extreme ambient conditions

  • Fewer process measuring points – multivariable measurement (flow, density, temperature)

  • Space-saving installation – no in/outlet run needs

  • Space-saving transmitter – full functionality on the smallest footprint

  • Fast commissioning – pre-configured devices

  • Automatic recovery of data for servicing

Application

  • Measuring principle operates independently of physical fluid properties such as viscosity or density

  • Accurate measurement of liquefied petroleum gas in refueling and distribution applications.

Device properties

  • Flow rates up to 70 000 kg/h (2570 lb/min)

  • Volume flow calculation according to API table 53

  • Robust, compact transmitter housing

  • Modbus RS485

  • Designed to meet application needs

Specification

Measuring PrincipleCoriolis
Product headlineThe refueling and distribution application flowmeter with easy sytem integration. Accurate measurement of liquefied petroleum gas in refueling and distribution applications.
Sensor featuresExcellent operational safety – reliable under extreme ambient conditions. Fewer process measuring points – multivariable measurement (flow, density, temperature). Space-saving installation – no in/outlet run needs. Flow rates up to 70 000 kg/h (2570 lb/min). Volume flow calculation according to API table 53.
Transmitter featuresSpace-saving transmitter – full functionality on the smallest footprint. Fast commissioning – pre-configured devices. Automatic recovery of data for servicing. Robust, compact transmitter housing. Modbus RS485.
Nominal diameter rangeDN 8 to 50 (³⁄₈ to 2")
Wetted materialsMeasuring tube: 1.4539 (904L) 
Connection: 1.4404 (316/316L)
Measured variablesMass flow, density, temperature, volume flow, corrected volume flow, reference density
Max. measured errorMass flow (liquid): ±0.20 % 
Volume flow (liquid): ±0.30 %
Measuring range0 to 70 000 kg/h (0 to 2570 lb/min)
Max. process pressurePN 40, Class 300
Medium temperature range–50 to +125 °C (–58 to +257 °F)
Ambient temperature range–40 to +60 °C (–40 to +140 °F)
Sensor housing material1.4301 (304), corrosion resistant
Transmitter housing materialAlSi10Mg, coated
Degree of protectionIP66/67, type 4X enclosure
Display/OperationNo local operation 
Configuration via operating tools possible
OutputsNone
InputsNone
Digital communicationModbus RS485
Power supplyDC 20 to 30 V
Hazardous area approvalsATEX, IECEx, cCSAus, NEPSI, INMETRO
Product safetyCE, C-Tick
Metrological approvals and certificatesCalibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)
Pressure approvals and certificatesPED, CRN
Material certificates3.1‐Material